Air-Sampler LKS30
German translation
Luftkeimsammler LKS30Category
SamplingSpecification
Round jet impactor for sampling air borne mikro-organisms on standard petri-dishes- Approved round-jet impation process
- Samples air- and gas borne bacteria and fungi on nutrient media in
- Standard-petri dishes
- To prove for micro-organisms that can be cultivated
- The positioning of the petri dishes allows sampling in all positions
- By connecting a hose samplings can also be done in cavities
- Identical air flow volume as in the particle sampler PS30 and the filter adpater FA 30 : 30 l/min
Shipment
air sampler, 2,5m hose 8mm x 2mm, manual, cleaning padsPrinciple
The technical specifications of the LKS 30:Air flow volume: 30 litres/min is recommended, it should not fall below 25 litres/min.
Sample volume: 90 to 100 litres in normal and slightly polluted environments 50 litres when pollution is higher.
Grade of separation: 0,9 µm aerodynamic diameter (dae50) at 30 litre/min air flow volume
Attitude: free due to the fixed positioning of the petri disch
Sample medium: standard petri dish, 90...95 mm x 15 mm (diameter x hight)
Negative pressure: ca. 0,1 bar at 30 Liter/min air flow volume.
Nozzle for sample air: can be extended with a hose of 30,5 mm (1,5 inch) internal diameter
Connection to pump: via nozzle for hose of 8 mm internal diameter
Size: 85 mm x 120 mm x 110mm (H x B xD)
Weight: 1. 080 g
Thread for tripod: UNC 1/4 inch (photo) and UNC 3/8 inch (microphone).
Note: subject to alterations
Handbook
Datasheet
Application
The air sampler LKS30 is ideally suited:- for microbiological air samplings in the health-care system and organic architecture.
- for microbiological examination of controlled- and clean air environments and filters.
- for microbiological examinations in the pharmaceutical-, cosmetic- and food industry.
Example and Functional principle